Commands listed in alphabetical order
Conditions
Requires optio
n 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:DELete “POWER<x>”
Examples
POWER:DELETE “POWER1” deletes the POWER1 measurement badge.
POWer:POWer<x>:AUTOSet (No Query Form)
This command executes power autoset for the specied power measurement
badge. The power measurement badge is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:AUTOSet EXECute
Examples
POWer:POWer1:AUTOSet EXECute executes the power autoset for the power
measurement badge Power 1.
POWer:POWer<x>:CYCLEAmp:INPUTSOurce
This command sets or queries the input source for cycle amplitude measurement
in the specied power measurement badge. The power measurement badge
is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:CYCLEAmp:IN PUTSOurce {CH<x> | MATH<x> | REF<x>}
POWer:POWer<x>:CYCLEAmp:IN PUTSOurce?
Arguments
CH<x> = A channel specier; <x> is 1 through 8 and is limited by the number
of FlexChannels in your instrument.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-529