Commands listed in alphabetical order
CH<x>:PROBEF
unc:EXTDBatten
This command sets or queries the input-output ratio (expressed in decibel u nits) of
external attenuation or gain between the signal and the instrument input channels.
The channel i
sspecied by x.
The query form of this command returns the user-specied attenuation in decibels.
Group
Vertical
Syntax
CH<x>:PROBEFunc:EXTDBatte n <NR3>
CH<x>:PROBEFunc:EXTDBatte n?
Related Commands
CH<x>:PROBEFunc:EXTAtten
Arguments
<NR3> is the a ttenuation value, which is specied in the range from -200.00 dB
to 200.00 dB.
Examples
CH3:PROBEFUNC:EXTDBATTEN 2.5 species an external 2.5 dB attenuator
on Channel 3.
CH1:PROBEFUNC:EXTDBATTEN? might return
:CH1:PROBEFUNC:EXTDBATTEE N 2.5000E+00, indicating that the attenuation
for Channel 1 is 2.5 dB.
CH<x>:PROBEFunc:EXTUnits
This command sets the unit of measurement for the external attenuator of the
specied channel. The channel is specied by x. The alternate u nits are used if
they are enabled. Use the CH<x>:PROBEFunc:EXTUnits:STATE command to
enable or disable the alternate units.
Group
Vertical
Syntax
CH<x>:PROBEFunc:EXTUnits <QString>
CH<x>:PROBEFunc:EXTUnits?
Related Commands
CH<x>:PRObe:UNIts?
Arguments
<QString> indicates the attenuation unit of measurement for the speci ed
channel.