EasyManuals Logo

Tektronix MSO58 User Manual

Tektronix MSO58
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #218 background imageLoading...
Page #218 background image
Commands listed in alphabetical order
CH<x>:PROBEF
unc:EXTDBatten
This command sets or queries the input-output ratio (expressed in decibel u nits) of
external attenuation or gain between the signal and the instrument input channels.
The channel i
sspecied by x.
The query form of this command returns the user-specied attenuation in decibels.
Group
Vertical
Syntax
CH<x>:PROBEFunc:EXTDBatte n <NR3>
CH<x>:PROBEFunc:EXTDBatte n?
Related Commands
CH<x>:PROBEFunc:EXTAtten
Arguments
<NR3> is the a ttenuation value, which is specied in the range from -200.00 dB
to 200.00 dB.
Examples
CH3:PROBEFUNC:EXTDBATTEN 2.5 species an external 2.5 dB attenuator
on Channel 3.
CH1:PROBEFUNC:EXTDBATTEN? might return
:CH1:PROBEFUNC:EXTDBATTEE N 2.5000E+00, indicating that the attenuation
for Channel 1 is 2.5 dB.
CH<x>:PROBEFunc:EXTUnits
This command sets the unit of measurement for the external attenuator of the
specied channel. The channel is specied by x. The alternate u nits are used if
they are enabled. Use the CH<x>:PROBEFunc:EXTUnits:STATE command to
enable or disable the alternate units.
Group
Vertical
Syntax
CH<x>:PROBEFunc:EXTUnits <QString>
CH<x>:PROBEFunc:EXTUnits?
Related Commands
CH<x>:PRObe:UNIts?
Arguments
<QString> indicates the attenuation unit of measurement for the speci ed
channel.
2-202 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals