Commands listed in alphabetical order
MEASUrement:
REFLevels:PERCent:TYPE
This command sets or queries the referen ce level percent type for the measurement.
Group
Measurement
Syntax
MEASUrement:REFLevels:PER Cent:TYPE
{TENNinety|TWENtyeight
y|CUSTom}
MEASUrement:REFLevels:PER Cent:TYPE?
Arguments
TENNinety sets the values for Low, Mid and High Ref to 10%, 50% and 90%
respectively.
TWENtyeighty sets the values for Low, Mid and High Ref are set to 20%, 50%
and 80% respectively.
CUSTom allows setting other reference level percents.
Examples
MEASUrement:REFLevels:PER Cent:TYPE TENNINETY sets the values for
Low, Mid and High Ref to 10%, 50% and 90% respectively.
MEASUrement:REFLevels:PER Cent:TYPE? might return
:MEASUREMENT:REFLEVELS:PE RCENT:TYPE CUSTOM indicating
that custom reference levels can be set.
MEASUrement:REFLevels:TYPE
This command sets or queries the shared reference level method used for sources
of measurement calculations.
Group
Measurement
Syntax
MEASUrement:REFLevels:TYP E {GLOBal|PERSource}
MEASUrement:REFLevels:TYP E?
Arguments
GLOBal shares reference levels across measurements.
PERSource causes reference levels to b e used on individual measurements.
Examples
MEASUREMENT:REFLEVELS:TYP E PERSource causes reference levels to be used
on individual measurements.
2-506 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals