Commands listed in alphabetical order
Examples
MEASUrement:M
EAS2:BITEnd 60
sets the bit end to 60% of the unit interval.
MEASU
rement:MEAS2:BITEnd?
might return :MEASUREME NT:MEAS2:BITEND
50.0000
indicating the bit end is 50%.
MEASUrement:MEAS<x>:BITPcnt
This command sets or queries the bit center as a percentage of the unit interval.
Measurements are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:BITPc nt <NR3>
MEASUrement:MEAS<x>:BITPc nt?
Arguments
<NR3> is the bit center percentage value to be measured for the bit type selected.
Examples
MEASUrement:MEAS2:BITPcnt 55 sets the bit center to 55% of the unit
interval.
MEASUrement:MEAS2:BITPcnt ? might return
:MEASUREMENT:MEAS2:BITPCN T 50.0000 indicating bit center is 50% of
the unit interval.
MEASUrement:MEAS<x>:BITSTart
This command sets or queries the bit start as a percentage of the unit interval.
Measurements are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:BITST art <NR3>
MEASUrement:MEAS<x>:BITST art?
Arguments
<NR3> is the bit start.
Examples
MEASUrement:MEAS2:BITSTar t 45 sets the bit start to 45% of the unit interval.
MEASUrement:MEAS2:BITSTar t? might return
:MEASUREMENT:MEAS2:BITSTA RT 50.0000 indicating the bit start is 50%
of the unit interval.
2-424 MSO54, MSO56, MSO58, MSO58LP Programmer