Commands listed in alphabetical order
Syntax
DISplay:PLOTV
iew<x>:CURSor:VBArs:DELTa?
Returns
The delta T cursor readout value is returned.
Examples
DISplay:PLOTView1:CURSor: VBArs:DELTa? might return
:DISPLAY:PLOTVIEW1:CURSOR :VBARS:DELTA 119.7593E-9 indicating
the
VBARS DE
LTA
is set to 119.76 ns.
DISplay:PLOTView<x>:CURSor:VBArs:UNIts? (Query Only)
This command queries the VBArs cursor readout units of the specied cursor
in the specied view.
Group
Cursor
Syntax
DISplay:PLOTView<x>:CURSo r:VBArs:UNIts?
Returns
The VBArs cursor readout units are returned.
Examples
DISplay:PLOTView<x>:CURSo r:VBArs:UNIts? might return
:DISPLAY:PLOTVIEW5:CURSOR :VBARS:UNITS "V" indicating the units are
Volts.
DISplay:PLOTView<x>:CURSor:WAVEform:APOSition
This command sets or queries the waveform cursor A horizontal position of the
specied cursor in the specied view.
NOTE. In case of XY plot, this command has no effect when used to set the value.
In case of bathtub plot, this command sets or returns the cursor A vertical position.
For all other plots, this command
sets or returns the cursor A horizontal position.
Group
Cursor
Syntax
DISplay:PLOTView<x>:CURSo r:WAVEform:APOSition <NR3 >
DISplay:PLOTView<x>:CURSo r:WAVEform:APOSition?
Arguments
<NR3> is the horizontal cursor A position.
2-258 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals