EasyManua.ls Logo

Tektronix MSO58 - TRIGger:{A|B}:PULSEWidth:SOUrce; TRIGger:{A|B}:PULSEWidth:WHEn

Tektronix MSO58
1026 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands listed in alphabetical order
Syntax
TRIGger:{A|B}
:PULSEWidth:POLarity {NEGative|POSiti ve}
TRIGger:{A|B}:PULSEWidth: POLarity?
Arguments
NEGative specifies a negative pulse.
POSitive specifies a positive pulse.
Examples
TRIGge
r:A:PULSEWidth:POLarity NEGA TIVE
sets the pulse polarity to
negative.
TRIGg
er:A:PULSEWidth:POLarity?
might return
:TRIGGER:A:WIDTH:POLARITY POSITIVE indicating a positive pulse.
TRIGger:{A|B}:PULSEWidth:SOUrce
This command specifies the source waveform for a pulse width trigger.
Group
Trigger
Syntax
TRIGger:{A|B}:PULSEWidth: SOUrce {CH<x>|CH<x>_D<y>}
TRIGger:{A|B}:PULSEWidth: SOUrce?
Arguments
CH<x> specifies an analog input channel as the pulse-width trigger source.
CH<x>_D,y> specifies an digital input channel as the pulse-width trigger source.
Examples
TRIGGER:A:PULSEWIDTH:SOUR CE CH1 sets channel 1 as the pulse width
source.
TRIGGER:A:PULSEWIDTH:SOUR CE? might r eturn
:TRIGGER:A:PULSEWIDTH:SOU RCE CH1 indicating that channel 1 is the
pulse width trigger source.
TRIGger:{A|B}:PULSEWidth:WHEn
This command specifies to trigger when a pulse is detected with a width (duration)
that is less than, greater than, equal to, or unequal to a specified value (set using
TRIGger:{A|B}:PULSEWidth:LOWLimit), OR whose width falls outside of or
within a specified range of two values (set using TRIGger:{A|B}:PULSEWidth:
LOWLimit and TRIGger:{A|B}:PULSEWidth:HIGHLimit).
2-858 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents