Commands listed in alphabetical order
Examples
BUS:B1:SPI:ID
LETime 0.0 00004
sets the idle time to 4 μs.
BUS:B1:SPI:IDLETime? might return :BU S:B1:SPI:IDLETIME 5.0000E -6,
indicating th
attheidletimeissetto5μs.
BUS:B<x>:SPI:SELect:POLarity
This command sets or queries the SPI Slave Select (SS) polarity for the specified
bus. The bus is specified by x.
Conditions
Requires option SR-EMBD.
Group
Bus
Syntax
BUS:B<x>:SPI:SELect:POLari ty {LOW|HIGH}
BUS:B<x>:SPI:SELect:POLari ty?
Related Commands
BUS:B<x>:SPI:SELect:SOUrce
Arguments
LOW sets an active low polarity.
HIGH sets an active high polarity.
Examples
BUS:B1:SPI:SELECT:POLA
RITY HI GH
sets the SPI Slave Select polarity for
the b us B1 to active high.
BUS:B1:SPI:SELECT:POL
ARITY?
might return
:BUS:B1:SPI:SELECT:POLARIT Y HIGH, indicating that the SPI Slave Select
polarity for the bus B1 is set to active high.
BUS:B<x>:SPI:SELect
:SOUrce
This co mmand sets or queries the SPI Slave Select (SS) s ource for the specified
bus. The bus is specified by x.
Conditions
Requires option SR-EMBD.
Group
Bus
MSO54, MSO56, MSO58, MSO58LP Programmer 2-175