Commands listed in alphabetical order
MEASUrement:
MEAS<x>:FILTers:HIGHPass:SPEC
This command sets or queries the high pass filter order for the measurement.
Measurements are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:FILTer s:HIGHPass:SPEC
{NONE|FIRST|SECOND|THIRD}
MEASUrement:MEAS<x>:FILTer s:HIGHPass:SPEC?
Arguments
NONE specifies no filter.
FIRST specifies a first-order filter.
SECOND specifies a second-order filter.
THIRD specifies a third-order filter.
Examples
MEASUrement:MEAS1:FILTers: HIGHPass:SPEC FIRST specifies a
first-order filter.
MEASUrement:MEAS1:FILTers: HIGHPass:SPEC? might return
:MEASUREMENT:MEAS1:FILTERS :HIGHPASS:SPEC NONE indicating no high
pass filter is used.
MEASUrement:MEAS<x>:FILTers:LOWPass:FREQ
This command sets or queries the low pass filter cutoff frequency for the
measurement. Measurements are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:FILTer s:LOWPass:FREQ <NR3>
MEASUrement:MEAS<x>:FILTer s:LOWPass:FREQ?
Arguments
<NR3> is the current low pass filter frequency.
Examples
MEASUrement:MEAS1:FILTers: LOWPass:FREQ 5.0E+6 sets the low pass
frequency to 5.0 MHz.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-449