EasyManua.ls Logo

Tektronix MSO58 - TRIGger:{A|B}:BUS:B<x>:FLEXray:FRAMEID:VALue; TRIGger:{A|B}:BUS:B<x>:FLEXray:FRAMEType

Tektronix MSO58
1026 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands listed in alphabetical order
Examples
TRIGGER:A:BUS
:B1:FLEXRAY:FRAMEID:QUALIFIE R LESSTHAN
sets the
frame ID qualifier to less than.
TRIGGER:A:BUS:B1:FLEXRAY: FRAMEID:QUALIFIER? might return
TRIGGER:A:BUS:B1:FLEXRAY: FRAMEID:QUALIFIER EQUAL indicating the
frame ID qualifier is set to equal.
TRIGger:{A|B}:BUS:B<x>:FLEXray:FRAMEID:VALue
This command specifies the low value when triggering on the FlexRay bus frame
ID field. (Use TRIGger:{A|B}:BUS:B<x>:FLEXray:FRAMEID:HIVALue to
set the high value.) The trigger condition needs to be set to IDentifier (using
TRIGger:{A|B}:BUS:B<x>:FLEXray:CONDition). The bus number is specified
by <x>.
Conditions
Requires the SR-AUTO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:FL EXray:FRAMEID:VALue <QStr ing>
TRIGger:{A|B}:BUS:B<x>:FL EXray:FRAMEID:VALue?
Arguments
<QString> is a quoted string that is the FlexRay frame ID low value.
Examples
TRIGGER:A:BUS:B1:FLEXRAY: FRAMEID:VALUE "11001100101" sets the
frame ID value to 11001100101.
TRIGGER:A:BUS:B1:FLEXRAY: FRAMEID:VALUE? might return
TRIGGER:A:BUS:B1:FLEXRAY: FRAMEID:VALUE "XXXXXXXXXXX" indicating
the frame ID value is don't care.
TRIGger:{A|B}:BUS:B<x>:FLEXray:FRAMEType
This command specifies the frame type (normal, payload, null, sync or startup)
when triggering on the FlexRay bus signal. The t rigger condition needs to be
set to FRAMEType (using TRIGger:{A|B}:BUS:B<x>:FLEXray:CONDition).
The bus number is specified by <x>.
Conditions
Requires the SR-AUTO Triggering and Analysis application.
Group
Trigger