Commands listed in alphabetical order
Group
Measurement
Syntax
MEASUrement:REF<x>:REFLeve ls:PERCent:RISELow <NR3>
MEASUrement:REF<x>:REFLeve ls:PERCent:RISELow?
Arguments
<NR3> is the percentage u sed to calculate the low reference level of the rising edge.
Examples
MEASUrement:REF1:REFLevels :PERCent:RISELow 5 sets the low reference
level to 5%.
MEASUrement:REF1:REFLevels :PERCent:RISELow? might return
:MEASUREMENT:REF1:REFLEVEL S:
PERCENT:RISELOW 10.0000
indicating
the reference level is 10%.
MEASUrement:REF<x>:REFLevels:PERCent:RISEMid
This command sets or queries the percentage (where 100% is equal to TOP and
0% is equal to BASE) used to calculate the mid reference level of the rising edge
when the measurement's ref level method is set to percent.
Group
Measurement
Syntax
MEASUrement:REF<x>:REFLeve ls:PERCent:RISEMid <NR3>
MEASUrement:REF<x>:REFLeve ls:PERCent:RISEMid?
Arguments
<NR3> is the percentage used to calculate the mid reference level of the rising
edge.
Examples
MEASUrement:REF1:REFLevels :PERCent:RISEMid 55 sets the mid
reference level to 55%.
MEASUrement:REF1:REFLevels :PERCent:RISEMid? might return
:MEASUREMENT:REF1:REFLEVEL S:PERCENT:RISEMID 50.0000 indicating
the mid reference level is 50%.
MEASUrement:REF<x>:REFLevels:PERCent:TYPE
This command sets or queries the reference level percent type for the measurement.
Group
Measurement
MSO54, MSO56, MSO58, MSO58LP Programmer 2-515