Commands listed in alphabetical order
Examples
DIAG:LOOP:OPT
ION ALWAYS
sets loop option to ALWAYS.
DIAG:LOOP:OPTION? returns current looping option value.
DIAg:LOOP:OPTion:NTIMes
This command sets or queries how many loops to run, if N-times is being used.
Group
Self Test
Syntax
DIAg:LOOP:OPTion:NTIMes <NR1>
DIAg:LOOP:OPTion:NTIMes?
Arguments
<NR1> is how many loops to run.
Examples
DIAG:LOOP:OPTION:NTIMES 2 sets diagnostics to loop 2 times.
DIAG:LOOP:OPTION:NTIMES? might return :DIAG:LOOP:OPTION:NTIMES 1
indicating diagnostics will run once.
DIAg:LOOP:STOP (No Query Form)
Request that diagnostics stop looping.
Group
Self Test
Syntax
DIAg:LOOP:STOP
Examples
DIAG:LOOP:STOP stops diagnostics looping.
DIAg:MODe
This command sets or queries the diagnostics mode.
Group
Self Test
Syntax
DIAg:MODe {POST|EXTENDED|SERVICE}
DIAg:MODe?
MSO54, MSO56, MSO58, MSO58LP Programmer 2-223