Commands listed in alphabetical order
Group
Measurement
Syntax
MEASUrement:MEAS<x>:FILTe rs:GLOBal {OFF|ON|0|1}
MEASUrement:MEAS<x>:FILTe rs:GLOBal?
Arguments
OFF causes filter settings to be changed independently for each individual
measurement.
ON applies global filter settings to all the measurements' filter settings.
0 causes filter settings to be changed independently for each individual
measurement.
1 applies global filter settings to all the measurements' filter settings.
Examples
MEASUrement:MEAS<x>:FILTe rs:GLOBal OFF causes filter s ettings to be
changed independently for each individual measurement.
MEASUrement:MEAS<x>:FILTe rs:GLOBal? might return
:MEASUREMENT:MEAS1:FILTER S:GLOBAL 1 indicating filter settings are for
all measurements.
MEASUrement:MEAS<x>:FILTers:HIGHPass:FREQ
This command sets or queries the high pass filter frequency for the measurement.
Measurements are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:FILTe rs:HIGHPass:FREQ <NR3>
MEASUrement:MEAS<x>:FILTe rs:HIGHPass:FREQ?
Arguments
<NR3> is the current high pass filter frequency.
Examples
MEASUrement:MEAS1:FILTers :HIGHPass:FREQ 15.0E+6 sets the high
pass frequency to 15.0 MHz.
MEASUrement:MEAS1:FILTers :HIGHPass:FREQ? might return
:MEASUREMENT:MEAS1:FILTER S:HIGHPASS:FREQ 10.0000E+6 indicating
the high pass frequency is set to 10.0 MHz.
2-448 MSO54, MSO56, MSO58, MSO58LP Programmer