EasyManua.ls Logo

Tektronix MSO58 - TRIGger:{A|B}:BUS:B<x>:I2 C:CONDition

Tektronix MSO58
1026 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands listed in alphabetical order
Conditions
Requires the SR
-EMBD Triggering a nd Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:I2C :ADDRess:VALue <QString>
TRIGger:{A|B}:BUS:B<x>:I2C :ADDRess:VALue?
Arguments
<QString> is up to 7 or 10-bits depending on the address mode that specifies the
address. The only allowed characters in the
QString are 0, 1, and X.
Examples
TRIGGER:A:BUS:B1:I2C:ADDRE SS:VALUE "1011" sets the I
2
C address value
to XXX1011.
TRIGGER:A:BUS:B1:I2C:ADDRE SS:VALUE? might return
:TRIGGER:A:BUS:B1:I2C:ADDR ESS:VALUE "XXXXXXX" indicating
the address value is set to XXXXXXX.
TRIGger:{A|B}:BUS:B<x>:I2C:CONDition
This command specifies the trigger condition for an I
2
C trigger. The bus number
is specified by <x>.
Conditions
Requires the SR-EMBD Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:I2C :CONDition
{STARt|STOP|REPEATstart|AC KMISS|ADDRess|DATa|ADDRANDDATA}
TRIGger:{A|B}:BUS:B<x>:I2C :CONDition?
Arguments
STARt specifies a search based on start condition.
STOP specifies a search based on stop condition.
REPEATstart specifies a search b ased on repeat of start condition.
ACKMISS specifies a search based on missing acknowledgement condition.
ADDRess specifies a search based on addres s.
DATa specifies a search based on data.
ADDRANDDATA specifies a search based on address and data.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-807