Commands listed in alphabetical order
Syntax
POWer:POWer<x
>:LINERIPPLE:INPUTSOurce
{CH<x> | MATH<x> |
REF<x>}
POWer:POWer<x>:LINERIPPLE: INPUTSOurce?
Arguments
CH<x> = A channel specier; <x> is 1 through 8 and is limited by the number
of FlexChannels in your instrument.
MATH<x> = A math waveform specier; <x> is 1.
REF<x> = A reference waveform specier; <x> is 1.
Examples
POWer:POWer1:LINERIPPLE:IN PUTSOurce CH1 sets channel 1 as the input
source for line r ipple measurement of power measurement badge Power 1.
POWer:POWer<x>:LINERIPPLE:LFREQuency
This command sets or queries the frequency present for line ripple measurement
in the specied power measurement badge. The power measurement badge
is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:LINERIPPLE: LFREQuency {FIFty | SIXty |
FOURHundred}
POWer:POWer<x>:LINERIPPLE: LFREQuency?
Example
s
POWer:POWer1:LINERIPPLE:LF REQuency FIFTY sets the frequency present
for line ripple measurement as 50 Hz.
POWer:POWer<x>:NDUTYCYCLE:EDGEType
This c
ommand sets or queries the clock edge type for negative duty cycle
measurement in the specied power measurement badge. The power measurement
badge is specied by x.
Conditions