Commands listed in alphabetical order
Arguments
INTERnal speci
fies the internal 10 MHz crystal oscillator as the time b ase
reference.
EXTernal spec
ifies the user-supplied external signal as the time base reference.
Examples
ROSC:SOURCE INTERNAL specifies the internal 10 MHz crystal oscillator as
the time bas
e reference.
ROSC:SOURCE? might return :R OSC:SOURCE INTERNAL, indicating that the
10 MHz crys
tal oscillator is being used as the time base reference.
ROSc:STATE? (Query Only)
This query-only command returns whether the time base reference oscillator is
locked. This command will return either LOCKED or UNLOCKED.
Group
Miscellaneous
Syntax
ROSc:STATE?
Returns
LOCK
ED
indicates the reference oscillator is locked.
UNLOCKED indicates the reference oscillator is not locked.
Examples
ROSC:STATE? might return :ROSC:STATE LOCKED, indicating that the time
base reference is locked.
*RST (No Query Form)
This command (no query form) resets the instrument to the factory default settings.
This command does the following:
Recalls the default instrument setup.
Clears the current *DDT command.
Disables aliases (:ALIAS:STATE 0).
Disables the user password (for the *PUD command).
MSO54, MSO56, MSO58, MSO58LP Programmer 2-615