Commands listed in alphabetical order
Arguments
<QString> is a q
uoted string that is the binary data string used for LIN trigger if
the trigger condition is ID or IDANDDATA.
Examples
TRIGGER:A:BUS:B1:LIN:DATA: HIVALUE "11001010" sets the high v alue
to
11001010.
TRIGGER:A:BUS:B1:LIN:DATA: HIVALUE? might return
TRIGGER:A:BUS:B1:LIN:DATA: HIVALUE "XXXXXXXX"indicating
the high value is don't care.
TRIGger:{A|B}:BUS:B<x>:LIN:DATa:QUALifier
This command specifies the LIN data qualifier. This only applies if the trigger
condition is IDANDDATA or DATA. The b us number is speci fi ed by <x>.
Conditions
Requires the SR-AUTO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:LIN :DATa:QUALifier
{LESSthan|MOREthan|EQual|U NEQual|LESSEQual|MOREEQual|
INrange|OUTrange}
TRIGger:{A|B}:BUS:B<x>:LIN :DATa:QUALifier?
Arguments
LESSthan sets the LIN data qualifier to less than.
MOREthan sets the LIN data qualifier to greater than.
EQual sets the LIN data qualifier to equal.
UNEQual sets the LIN data qualifier to not equal.
LESSEQual sets the LIN data qualifier to less than or equal.
MOREEQual sets the LIN data qualifier to greater than or equal.
INrange sets the LIN data qualifier to in range.
OUTrange sets the LIN data qualifier to out of range.
Examples
TRIGGER:A:BUS:B1:LIN:DATA: QUALIFIER OUTRANGE sets the data qualifier
to out of range.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-811