Commands listed in alphabetical order
Arguments
<QString> is a q
uoted string that is the binary data string used for LIN trigger if
the trigger condition is ID or IDANDDATA.
Examples
TRIGGER:A:BUS:B1:LIN:DATA: HIVALUE "11001010" sets the high v alue
to
11001010.
TRIGGER:A:BUS:B1:LIN:DATA: HIVALUE? might return
TRIGGER:A:BUS:B1:LIN:DATA: HIVALUE "XXXXXXXX"indicating
the high value is don't care.
TRIGger:{A|B}:BUS:B<x>:LIN:DATa:QUALier
This command species the LIN data qualier. This only applies if the trigger
condition is IDANDDATA or DATA. The b us number is speci ed by <x>.
Conditions
Requires the SR-AUTO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:LIN :DATa:QUALifier
{LESSthan|MOREthan|EQual|U NEQual|LESSEQual|MOREEQual|
INrange|OUTrange}
TRIGger:{A|B}:BUS:B<x>:LIN :DATa:QUALifier?
Arguments
LESSthan sets the LIN data qualier to less than.
MOREthan sets the LIN data qualier to greater than.
EQual sets the LIN data qualier to equal.
UNEQual sets the LIN data qualier to not equal.
LESSEQual sets the LIN data qualier to less than or equal.
MOREEQual sets the LIN data qualier to greater than or equal.
INrange sets the LIN data qualier to in range.
OUTrange sets the LIN data qualier to out of range.
Examples
TRIGGER:A:BUS:B1:LIN:DATA: QUALIFIER OUTRANGE sets the data qualier
to out of range.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-811

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals