Commands listed in alphabetical order
Examples
MEASUREMENT:F
ILTERS:BLANKINGTIME
3.0 sets the blanking time to 3.0.
MEASUREMENT:FILTERS:BLANKI NGTIME? might return
:MEASUREMENT
:FILTERS:BLANKINGTIME 4.000
0 indicating the
blanking time is 4.0.
MEASUrement:FILTers:HIGHPass:FREQ
This command sets or queries the global high pass filter frequency for the
measurement.
Group
Measurement
Syntax
MEASUrement:FILTers:HIGHPa ss:FREQ <NR3>
MEASUrement:FILTers:HIGHPa ss:FREQ?
Arguments
<NR3> is the current high pass filter frequency.
Examples
MEASUREMENT:FILTERS:HIGHPA SS:FREQ 20.0e6 sets the frequency to 20
MHz.
MEASUREMENT:FILTERS:HIGHPA SS:FREQ? might return
:MEASUREMENT:FILTERS:HIGHP ASS:FREQ 10.0000E+6 indicating
the filter frequency is set to 10.0 MHz.
MEASUrement:FILTers:HIGHPass:SPEC
This command sets or qu
eries the global high pass filter order for the measurement.
Group
Measurement
Syntax
MEASUrement:FILTers:HIGHPa ss:SPEC {NONE|FIRST|SECON D|THIRD}
MEASUrement:FILTers:HIGHPa ss:SPEC?
Arguments
NONE specifies no high pass filter.
FIRST specifies a first order high pass filter.
SECOND specifies a second order high pass filter.
THIRD specifies a third order high pass filter.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-403