Commands listed in alphabetical order
Syntax
MEASUrement:R
EF<x>:REFLevels:PERCent:TYPE
{TENNinety|TWENtyeighty|C USTom}
MEASUrement:REF<x>:REFLev els:PERCent:TYPE?
Arguments
TENNinety sets the values for Low, Mid and High Ref to 10%, 50% and 90%
respectively.
TWENtyeighty sets the values for Low, Mid and High Ref are set to 20%, 50%
and 80% respectively.
CUSTom allows setting other reference level percents.
Examples
MEASUrement:REF1:REFLevel s:PERCent:TYPE TWENTYEIGHTY sets the
values for Low, Mid and High Ref are set to 20%, 50% and 80% respectively.
MEASUrement:REF1:REFLevel s:PERCent:TYPE? might return
:MEASUREMENT:REF1:REFLEVE LS:PERCENT:TYPE TENNINETY indicating the
values for Low, Mid and High Ref to 10%, 50% and 90% respectively.
MEASUrement:STATIstics:CYCLEMode
This command turns on and off cycle to cycle measurement statistics tracking
and affects computation and display of cycle-cycle statistics in the Measurement
Result table. It affects measurement sta tistics after being enabled and after new
data is acquired and measured.
Group
Measurement
Syntax
MEASUrement:STATIstics:CY CLEMode {OFF|ON|0|1}
MEASUrement:STATIstics:CY CLEMode?
Arguments
OFF turns off statistics for all measurements. This is the default value.
ON turns on statistics and displays all statistics for each measurement.
0 turns off statistics for all measurements.
1 turns on statistics and displays all s tatistics for each measurement.
Examples
MEASUREMENT:STATISTICS:CY CLEMODE OFF turns off statistics for all
measurements.
MEASUREMENT:STATISTICS:CY CLEMODE? might return
:MEASUREMENT:STATISTICS:C YCLEMODE 1 indicating that
statistics are displaye dforeachmeasurement.
2-516 MSO54, MSO56, MSO58, MSO58LP Programmer