Commands listed in alphabetical order
For Cycle Base m
easurement, the parameters is “LOW”.
For Cycle Max measurement, the parameters is “MAX”.
For Cycle Min measurement, the parameters is “MIN”.
Examples
POWer:POWer1:RESUlts:CURR entacq:STDDev? “TONLoss” might return
9.91E+37, indicating the standard deviation value of Ton Energy for the current
acquisition.
POWer:POWer<x>:RESUlts:CURRentacq:THDF? (Query Only)
This command queries the total harmonic distortion (fundamental) value for the
specied power measurement badge. The power measurement badge is specied
by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:RESUlts:CU RRentacq:THDF? "harmonics"
Examples
POWer:POWer1:RESUlts:CURR entacq:THDF? "harmonics" might return
96.9846996670887E-3, indicating the value of total harmonic distortion
(fundamental) for the power measurement badge 1.
POWer:POWer<x>:RESUlts:CURRentacq:THDR? (Query Only)
This command queries the total harmonic distortion (RMS) value for the specied
power measurement badge. The power measurement badge is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:RESUlts:CU RRentacq:THDR? "harmonics"