EasyManua.ls Logo

Tektronix MSO58 - MEASUrement:MEAS<x>:CLOCKRecovery:METHod; MEASUrement:MEAS<x>:CLOCKRecovery:MODel

Tektronix MSO58
1026 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands listed in alphabetical order
Syntax
MEASUrement:M
EAS<x>:CLOCKRecovery:MEANAUT OCalculate
{FIRST|EVERY}
MEASUrement:MEAS<x>:CLOCKR ecovery:MEANAUTOCalculate?
Arguments
FIRST calculates the clock on the first acquisition.
EVERY calculates the clock on every acquisition.
Examples
MEASUrement:MEAS1:CLOCKRec overy:MEANAUTOCalculate E VERY causes
the clock to be calculated on every acquisition.
MEASUrement:MEAS1:CLOCKRec overy:MEANAUTOCalculate? might return
:MEASUREMENT:MEAS1:CLOCKRE COVERY:MEANAUTOCALCULATE FIRST
indicating the clock is calculated on the first acquisition.
MEASUrement:MEAS<x>:CLOCKRecovery:METHod
This command sets or queries the clock recovery method for the measurement.
Measurements are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:CLOCKR ecovery:METHod
{PLL|CONSTANTCLOCK|EXPLICI TCLOCK}
MEASUrement:MEAS<x>:CLOCKR ecovery:METHod?
Arguments
PLL specifies a phase locked loop clock recovery method.
CONSTANTCLOCK specifies using a constant clock.
EXPLICITCLOCK specifies using an explicit clock.
Examples
MEASUrement:MEAS1:CLOCKRec overy:METHod PLL specifies a phase locked
loop clock recovery method.
MEASUrement:MEAS1:CLOCKRec overy:METHod? might return
:MEASUREMENT:MEAS1:CLOCKRE COVERY:METHOD CONSTANTCLOCK
indicating a constant clock is in u se.
MEASUrement:MEAS<x>:CLOCKRecovery:MODel
This command s ets or que ries the PLL clock recovery model used w h en PLL
clock recovery is used for the measurement. Measureme nts are specified by x.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-437