Commands listed in alphabetical order
Syntax
MATH:MATH<x>:
SPECTral:SUPPress {OFF|ON|0|1}
MATH:MATH<x>:SPECTral:SUPP ress?
Related Commands
MATH:MATH<x>:SPECTral:TYPE
Arguments
0 disables suppression threshold for the specied math waveform.
1 enables suppression threshold for the specied math waveform.
ON enables suppression threshold for the specied math waveform.
OFF disables suppression threshold for the specied math waveform.
Examples
MATH:MATH3:SPECTRAL:SUPPRE SS ON enables the suppression threshold on
the Math 3 spectral waveform.
MATH:MATH2:SPECTRAL:SUPPRE SS? might return
:MATH:MATH2:SPECTRAL:SUPPR ESS 0 indicating the suppression threshold is
disabled on the Math 2 spectral waveform.
MATH:MAT H<x>:SPECTral:SUPPress:VALue
This command sets or queries in volts the value of suppression threshold of the
specied math waveform. This requires the Spectral type to be Phase and the
Suppression to be enabled for this PI command to have any affect. The math
waveform is specied by x.
Group
Math
Syntax
MATH:MATH<x>:SPECTral:SUPP ress:VALue <NR3>
MATH:MATH<x>:SPECTral:SUPP ress:VALue?
Related Commands
MATH:MATH<x>:SPECTral:TYPE, MATH:MATH<x>:SPECTral:SUPPress
Arguments
<NR3> is the value of suppression threshold of the specied math waveform in
volts.
Examples
MATH:MATH3:SPECTRAL:SUPPRE SS:VALUE 100.0E-3 sets the suppression
thresholdofMath3to100mV.
MATH:MATH1:SPECTRAL:SUPPRE SS:VALUE? might return
MATH:MATH1:SPECTRAL:SUPPRE SS:VALUE 10.0000E+3 indicating
the suppression threshold of Math 1 is currently set to 10 kV.