Commands listed in alphabetical order
MEASUrement:
REFLevels:MODE
This command sets or queries how often reference levels are calculated.
Group
Measurement
Syntax
MEASUrement:REFLevels:MOD E {LATCh|CONTinuous}
MEASUrement:REFLevels:MOD E?
Arguments
LATCh calculate s reference levels only on the first acquisition after a statistics
reset.
CONTinuous calculates reference levels on every acquisition.
Examples
MEASUREMENT:REFLEVELS:MOD E CONTINUOUS calculates reference levels on
every acquisition.
MEASUREMENT:REFLEVELS:MOD E? might return
:MEASUREMENT:REFLEVELS:MO DE LATCH indicating reference levels are
calculated only on the first acquisition after a statistics reset.
MEASUrement:REFLevels:PERCent:FALLHigh
This command sets or queries the percentage (where 100% is equal to TOP and
0% is equal to BASE) used to calculate the high reference level of the falling edge
when the measurement's ref level method is set to percent.
Group
Measurement
Syntax
MEASUrement:REFLevels:PER Cent:FALLHigh <NR3>
MEASUrement:REFLevels:PER Cent:FALLHigh?
Arguments
<NR3> is the percentage used to calculate the high reference level of the falling
edge
Examples
MEASUrement:REFLevels:PER Cent:FALLHigh 90 sets the h igh reference
level to 90%.
MEASUrement:REFLevels:PER Cent:FALLHigh? might return
:MEASUREMENT:REFLEVELS:PE RCENT:FALLHIGH 80.0000 indicating the
high reference level is 80%.
2-502 MSO54, MSO56, MSO58, MSO58LP Programmer