Commands listed in alphabetical order
Examples
:TRIGger:AUXL
evel 1.2
sets trigger level for the Auxiliary Input to 1.2 Volts.
TRIGger:B:BY
This command selects or returns whether the B trigger occurs after a spec ified
number of events or a specified period of time after the A trigger.
Group
Trigger
Syntax
TRIGger:B:BY {EVENTS|TIMe}
TRIGger:B:BY?
Related Commands
TRIGger:B:EVENTS:COUNt, TRIGger:B:TIMe, TRIGger:B:STATE
Arguments
EVENTS sets the B trigger to take place following a set number of trigger
events after the A trigger occurs. The number of events is specified by
TRIGger:B:EVENTS:COUNt.
TIMe sets the B trigger to occur a set time after the A trigger event. The time
period is specified by
TRIGger:B:TIMe.
Examples
TRIGGER:B:BY TI ME sets the B trigger to occur at a set time after the A trigger
event.
TRIGGER:B:BY? might return :TRIGGER:B:BY EVENTS, indicating that the B
trigger takes place following a set number of trigger events after the A trigger
occurs.
TRIGger:B:EVENTS:COUNt
This command sets or queries the number of events that must occur before the
B trigger. The B trigger event count applies only if TRIGger:B:BY is set to
EVENTS.
Group
Trigger
Syntax
TRIGger:B:EVENTS:COUNt <NR1>
TRIGger:B:EVENTS:COUNt?
Related Commands
TRIGger:B:STAT E
2-878 MSO54, MSO56, MSO58, MSO58LP Programmer