Commands listed in alphabetical order
Returns
The total numbe
r of hours the instrument has been turned on since the NV
memory was last programme d.
Examples
TOTALUPTIME? might return :TOTALUPTIME 756 indicating the up time is
756 minutes.
TOUCHSCReen:CALibrate (No Query Form)
This command launches the touchscreen calibration procedure. This command
is equivalent to tapping the Calibrate Touchscreen control in the Utility->Self
Test menu.
Group
Self Test
Syntax
TOUCHSCReen:CALibrate STARt
Arguments
STARt launches the touchscreen calibration procedure.
Examples
TOUCHSCREEN:CALIBRATE START launches the touchscreen calibration
procedure.
TOUCHSCReen:STATe
This sets or queries the enabled state of the touch screen. This command is
equivalent to pushing the Touch Off button on the front panel.
To completely disable front panel operation, use the following two commands:
LOCK AL L;:TOUCHSCREEN:S TATE OFF. To re-enable the front panel, send
these two commands:
LOCK NO NE;:TOUCHSCREEN: STATE ON. The commands
must be sent in that order.
Group
Miscellaneous
Syntax
TOUCHSCReen:STATe {0|1|OFF|ON}
TOUCHSCReen:STATe?
Related Commands
LOCk
MSO54, MSO56, MSO58, MSO58LP Programmer 2-757

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals