EasyManua.ls Logo

Tektronix MSO58

Tektronix MSO58
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands listed in alphabetical order
Group
Display Contro
l
Syntax
DISplay:WAVEView<y>:REF<x >_DALL:FRAMe <NR1>
DISplay:WAVEView<y>:REF<x >_DALL:FRAMe?
Arguments
<NR1> is the selected frame of the specied digital ref.
Examples
DISplay:WAVEView<y>:REF<x >_DALL:FRAMe? might return
:DISPLAY:WAVEVIEW1:REF1_D ALL:FRAME 1, indicating the selected frame
is 1.
DVM (No Query Form)
Resets the D igital Voltmeter measurements and history.
Conditions
Requires DVM option.
Group
DVM
Syntax
DVM RES ET
Arguments
RESET species resetting DVM measurements and history.
Examples
DVM RES ET resets the DVM measurement and history.
DVM:AUTORange
Sets (or queries) the autorange state for the Digital Voltmeter.
NOTE. the DVM will not autorange as long as the DVM source is the same
channel as the trigger source.
Group
DVM
Syntax
DVM:AUTORange {0|1|OFF|ON}
DVM:AUTORange?
2-316 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO58

Related product manuals