Commands listed in alphabetical order
Examples
MEASUrement:M
EAS1:CCRESUlts:CURRentacq:ST DDev?
might return
:MEASUREMENT:MEAS1:CCRESUL TS:CURRENTACQ:STDDEV 9.91 E+37.
MEASUrement:MEAS<x>:CLOCKRecovery:ADVanced:METHod
This command sets or queries the advanced clock recovery method when advanced
clock recovery is used for the measurement. Measureme nts are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:CLOCKR ecovery:ADVanced:METHod
{NONE|NOMinal|PATTern}
MEASUrement:MEAS<x>:CLOCKR ecovery:ADVanced:METHod?
Arguments
NOMinal sets the advanced CRD method to Nominal Data Rate.
PATTern sets the advanced CDR method to use a Known Data Pattern.
NONE sets no advanced CRD method.
Examples
MEASUrement:MEAS1:CLOCKRec o
very:ADVanced:METHod PATTERN
sets
the advanced CDR method to use a Known Data Pattern.
MEASUrement:MEAS1:CLOCKRec overy:ADVanced:METHod? might
return
:MEASUREMENT:MEAS1:CLOCKREC OVERY:ADVANCED:METHOD NON E
indicating no advanced CRD method.
MEASUrement:MEAS<x>:CLOCKRecovery:CLOCKFrequency
This command sets or queries the clock frequency used when fixed constant clock
recovery is used for the measurement. Measurements are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:CLOCKR ecovery:CLOCKFrequency <N R3>
MEASUrement:MEAS<x>:CLOCKR ecovery:CLOCKFrequency?
Arguments
<NR3> is the clock frequency.
Examples
MEASUrement:MEAS1:CLOCKRec overy:CLOCKFrequency 2.0E +9 sets the
frequency to 2.0 GHz.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-431