EasyManua.ls Logo

Tektronix MSO58

Tektronix MSO58
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands listed in alphabetical order
Syntax
MEASUrement:M
EAS<x>:RESUlts:CURRentacq:ST DDev?
Returns
The standard deviation of values accumulated for the specied measurement
since the last statistical reset.
Examples
MEASUrement:MEAS<x>:RESUlt s:CURRentacq:STDDev? might return
:MEASUREMENT:MEAS1:RESULTS :CURRENTACQ:STDDEV 0.0E+0 indicating
the standard devi
ation for the measurement is 0.0 V.
MEASUrement:MEAS<x>:SIGNALType
This command sets or queries the signal type of source 1 for the measurement.
The measure ment number is specied by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SIGNAL Type {CLOCK|DATA|AUTO}
MEASUrement:MEAS<x>:SIGNAL Type?
Arguments
CLOCK species a clock signal type.
DATA species a data signal type.
AUTO automatically selects the signal type.
Examples
MEASUrement:MEAS1:SIGNALTy pe CLOCK species a clock signal type.
MEASUrement:MEAS1:SIGNALTy pe? might return
:MEASUREMENT:MEAS1:SIGNALT YPE AUTO indicating the instrument
automatically selects the signal type.
MEASUrement:MEAS<x>:SOUrce<x>
This command sets or queries the measurement source. The measurement number
and s ource a re specied by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce <x>
MSO54, MSO56, MSO58, MSO58LP Programmer 2-485