EasyManuals Logo
Home>Tektronix>Test Equipment>MSO58

Tektronix MSO58 User Manual

Tektronix MSO58
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #437 background imageLoading...
Page #437 background image
Commands listed in alphabetical order
Arguments
<NR3> is the min
imum measurement range limit value.
Examples
MEASUREMENT:MEASRANGE:MIN 30.0E-3 sets the minimum measurement range
limit to 30.0 mV.
MEASUREMENT:MEASRANGE:MIN? might return
:MEASUREMENT:MEASRANGE:MIN 0.0E+0 indicating minimum range limit is
0.0 V.
MEASUrement:MEASRange:STATE
This command sets or q ueries the global range state.
Group
Measurement
Syntax
MEASUrement:MEASRange:STAT E {OFF|ON|0|1}
MEASUrement:MEASRange:STAT E?
Arguments
OFF species that the measurement results are not limited.
ON species that the measurement results are limited to results with values
between the range minimum and maximum.
0 species that the measurement results are not limited.
1 species that the measurement results are limited to results with values between
the range minimum and maximum.
Examples
MEASUREMENT:MEASRANGE:STAT E 1 species that the measurement results are
limited to results with values between the range minimum and maximum.
MEASUREMENT:MEASRANGE:STAT E? might return
:MEASUREMENT:MEASRANGE:STA TE 0 indicating that the measurement results
are not limited.
MEASUrement:MEAS<x>:BER
This command sets or queries BER value for the measurement. Measurements
are specied by x.
Group
Measurement
MSO54, MSO56, MSO58, MSO58LP Programmer 2-421

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals