Commands listed in alphabetical order
TRIGger:{A|B
}:BUS:B<x>:CAN:DATa:SIZe
This command sets the length of the data string, in bytes, to be used when
triggering on a CAN bus signal. The trigger condition must be set to
IDANDDATA or
DATa (using TRIGger:{A|B}:BUS:B<x>:CAN:CONDition).
The bus number is specied by <x>.
Conditions
Requires the SR-AUTO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:CA N:DATa:SIZe <NR1>
TRIGger:{A|B}:BUS:B<x>:CA N:DATa:SIZe?
Arguments
<NR1> is the length of the data string in bytes.
Examples
TRIGger:A:BUS:B1:CAN:DATa :SIZe 2 sets the data size to 2 bytes.
TRIGger:A:BUS:B1:CAN:DATa :SIZe? might return
:TRIGGER:A:BUS:B1:CAN:DAT A:SIZE 1 indicating the data size is
setto1byte.
TRIGger:{A|B}:BUS:B<x>:CAN:DATa:VALue
This command sets the binary data value to be used when triggering on a CAN
bus signal. The trigger condition must be set to IDANDDATA or DATa (using
TRIGger:{A|B}:BUS:B<x>:CAN:CONDition). The bus number is specied
by <x>.
Conditions
Requires the SR-AUTO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:CA N:DATa:VALue <QString>
TRIGger:{A|B}:BUS:B<x>:CA N:DATa:VALue?
Arguments
<QString> is the data value in binary format. The only allowed c haracters in
the
QString are0,1,andX.