Screen element Description
Time remaining Time remaining until all injections for the plate are completed. Changes to Estimated
end time if the injection order for the injection or the plate is moved to a later position
in the run queue. See “Use the Run queue to monitor a run on the instrument” on
page 160.
Sample QC (or Spectral
QC or Install QC) for the
plate
See “Sample QC and quality alerts” on page 172.
Plate name Plate name.
User name Name of the user who was signed in when the run was started.
Plate position status and plate run status
Status
Description
Plate position status
Occupied The corresponding position in the instrument drawer contains a plate. These
positions are labeled as Link plate file in the instrument home screen.
Available The corresponding position in the instrument drawer is empty.
Plate run status
In queue An injection is waiting to start. The injection will start based on the order of the
injection list displayed in the Run Queue screen.
Running Data collection for the injection is in process.
Analyzing Data collection for the injection is complete, analysis is in process.
Exporting Result files are being sent to the Save location in the plate file.
Completed Analysis and export is complete, results are available.
Aborting/aborted An injection with status of Running was cancelled.
Cancelled An injection with status of In queue was cancelled.
Paused The instrument was paused and the "pause immediately" option was selected.
The pause and cancel commands can be issued in the SeqStudio
™
Flex Remote
Monitoring software or on the instrument.
Runs must be resumed on the instrument.
Chapter 1 Product information
About the SeqStudio
™
Flex Series Instrument Software v1.0
1
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
39