3.
Remove and retain the ABC. Insert an empty ABC to capture waste.
Figure 48 Location of pump, ABC, and pouch
1
Polymer pump block
2
Anode Buer Container (ABC), remove and replace with an empty ABC to capture waste
3
Pouch, remove polymer and replace with conditioning reagent
4.
Remove the polymer pouch by pr
essing the polymer pouch lever down.
Figure 49 Pouch slot and lever
1
Slot of the lever assembly
2
Label facing the instrument
3
Lever
Chapter 13 Maintain the Instrument
Maintenance wizard procedures
13
400
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide