Table 11 Summary tab—sequencing install run (continued)
Result Description
Spectral calibration
Spectral calibration status Pass or fail.
Spectral calibration
passing criteria
The number of capillaries that can be borrowed: 2 for 24-capillary instruments and
1 for 8-capillary instruments. For more information, see “Spectral calibration sharing
between capillaries” on page 321 and “Spectral calibration with borrowing enabled”
on page 322.
# caps passed/# caps
borrowed
The number of capillaries that met the spectral QV and condition number limits. See
“Spectral QV (quality value) and Condition number definitions and limits” on page 320.
The number of capillaries that were borrowed.
Capillaries tab—sequencing install run
Table 12 Capillaries tab—sequencing install run
Result Description
QV20 CRL (range) The QV20 CRL result and the base number range in which the QV20 CRL was
det
ermined.
QV20 basecall
(This result is referred to
as QV20 on the Details
screen.)
The percent accuracy of the basecalls in the sample sequence within the QV20
CRL range relative to the known reference sequence of the sequencing install
standard.
Chapter 12 Run calibrations and install checks
Perform an install run
12
SeqStudio
™
Flex Series Genetic Analyzer with Instrument Software v1.0 User Guide
341