EasyManua.ls Logo

Keithley 4200A-SCS - Page 1003

Keithley 4200A-SCS
1381 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section
13: LPT library function reference Model 4200A-SCS Parameter Analyzer
Reference Manual
13-158 4200A-901-01 Rev. C / February 2017
The pulse voltage high setting takes effect immediately during continuous pulse output. Otherwise,
the voltage high setting takes effect when the next trigger is initiated. The pulse_trig command is
used to trigger continuous or burst output.
The pulse_vlow, pulse_vhigh, and pulse_dc_output commands set the voltage value
output by the pulse channel when it is turned on (using pulse_output). If the output is already
enabled, these commands change the voltage level immediately, before the pulsing is started with a
pulse_trig command.
Example
pulse_vhigh(VPU1, 1, 2.5)
Sets the pulse voltage high value for channel 1 of the pulse card to 2.5 V.
Also see
pulse_dc_output (on page 13-140)
pulse_output (on page 13-146)
pulse_range (on page 13-149)
pulse_trig (on page 13-152)
pulse_vlow (on page 13-159)

Table of Contents