-SCS Parameter Analyzer Reference Manual Section 13:
LPT library function reference
4200A-901-01 Rev. C / February 2017 13-75
limitX
This command allows the programmer to specify a current or voltage limit other than the default limit of the
instrument.
Usage
int limiti(int instr_id, double limit_val);
int limitv(int instr_id, double limit_val);
instr_id
The instrument identification code of the instrument on which to impose a source
value limit
limit_val
The maximum level of the current or voltage; see Details
Details
The parameter limit_val is bidirectional. For example, the command limitv(SMU1, 10.0)
limits the voltage of the current source SMU1 to ±10.0 V. The command limiti(SMU1, 1.5E-3)
limits the current of the voltage source SMU1 to ±1.5 mA.
Use the limiti command to limit the current of a voltage source. Use the limitv command to limit
the voltage of a current source.
If the instrument is ranged below the programmed limit value, the instrument will temporarily limit to
full scale of range.
This command must be called in the test sequence before the associated forceX, pulseX,
bsweepX, sweepX, or searchX command is used to generate the voltage or current. The limitX
command also sets the top measurement range of an autoranged measurement.
The limits set within a particular test sequence are cleared when the devint or execut command is
called.
If you need a voltage limit greater than 20 V at a source-measure unit (SMU) that has been set to
force zero current, call the measv command to set the SMU to autorange to a higher range, or use
the rangev command to set a higher voltage range. Similarly, if you need a current limit of greater
than 10 mA at a SMU that has been set to force zero volts, call the measi command to set the SMU
to autorange to a higher range or use the rangev command to set a higher current range.

Table of Contents

Other manuals for Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200A-SCS and is the answer not in the manual?

Keithley 4200A-SCS Specifications

General IconGeneral
BrandKeithley
Model4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals