EasyManuals Logo
Home>Keithley>Measuring Instruments>4200A-SCS

Keithley 4200A-SCS Reference Manual

Keithley 4200A-SCS
1381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #770 background imageLoading...
Page #770 background image
Model 4200A
-SCS Parameter Analyzer Reference Manual Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017 9-59
KXCI pulse generator commands
The KXCI commands to control pulse generators are documented in the following topics.
The following documentation includes the corresponding LPT library functions for each KXCI
command string. Refer to LPT functions for PGUs and PMUs (on page 13-91) for details on pulsing
functions.
PD
This command sets the output impedance (pulse load).
Usage
PD A, BBB.BBBB
A
Pulse card channel number: 1 to 8; the largest value is the number of channels in
the system
BBB.BBBB
Pulse load, in ohms: 1.0 to 10e6; default 50.0
Corresponding LPT library function
pulse_load (on page 13-145)
Details
You can use this command to set the DUT impedance of the load (DUT) independently for each
channel from 1 Ω to 10 MΩ.
Example
PD 1, 1e3
This command string sets channel 1 of pulse card 1 for an output impedance of 1 kΩ.
Also see
None

Table of Contents

Other manuals for Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200A-SCS and is the answer not in the manual?

Keithley 4200A-SCS Specifications

General IconGeneral
BrandKeithley
Model4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals