REGFITXINT Formulator function
Performs a linear regression fit.
Usage
REGFITXINT(VX, VY, STARTPOS, ENDPOS)
VX
The name of any column (vector) listed under Columns
VY
The name of any column (vector) listed under Columns
STARTPOS
For the range of X and Y values to be fitted, the row number (index) of the starting
values
ENDPOS
For the range of X and Y values to be fitted, the row number (index) of the ending
values
Details
Performs a linear regression fit as follows:
• Fits the following relationship, of the form Y = a + bX, to a specified range of values in two
columns (vectors) (column VX containing X values and column VY containing Y values):
Y = REGFITYINT + REGFITSLP * X
where REGFITSLP and REGFITYINT are slope and Y-intercept constants.
• Returns the value of the X intercept for relationship above.
(–REGFITYINT/REGFITSLP).
If a VX or VY value at either STARTPOS or ENDPOS is an invalid number (that is, the value is #REF),
the function will not return a valid result.
Example
EARLYV = REGFITXINT(COLLECTORV, COLLECTORI, 25, LASTPOS(COLLECTORV))
Also see
REGFIT (on page 6-260)
REGFITYINT (on page 6-263)

Table of Contents

Other manuals for Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200A-SCS and is the answer not in the manual?

Keithley 4200A-SCS Specifications

General IconGeneral
BrandKeithley
Model4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals