13: LPT library function reference  Model 4200A-SCS Parameter Analyzer 
 
13-36  4200A-901-01 Rev. C / February 2017 
 
rtfary 
This command returns the force array determined by the instrument action. 
Usage 
int rtfary(double *results); 
 
The floating point array where the force values are stored 
Details 
This command eliminates the need to calculate the forced array in the application. 
When this command is used with one of the sweep routines, you can determine the exact forced 
value for each point in the sweep. 
When the test sequence is executed, the sweep command initiates the first step of the voltage or 
current sweep. The sweep then logs the force point that the buffer specified by the rtfary 
command. 
Locate the rtfary command before the sweep. The number of data points returned by the rtfary 
command is determined by the number of force points generated by the sweep. 
Example 
Refer to the examples for the smeasX and sweepX commands. 
Also see 
smeasX (on page 13-45) 
sweepX (on page 13-88) 
 
savgX 
This command makes an averaging measurement for every point in a sweep. 
Usage 
int savgi(int instr_id, double *result, long count, double delay); 
int savgv(int instr_id, double *result, long count, double delay); 
 
The instrument identification code of the measuring instrument 
The floating point array where the results are stored 
The number of measurements made at each point before the average is computed 
The time delay in seconds between each measurement within a given ramp step