13: LPT library function reference Model 4200A-SCS Parameter Analyzer
13-70 4200A-901-01 Rev. C / February 2017
Example
.
.
conpin(2, GND, 0);
conpin(SMU1, 1, 0);
limiti(SMU1, 2.0E-4); /* Limit 1 mA to 200 uA. */
forcev(SMU1, 40.0); /* Apply 40.0 V. */
measi(SMU1, &ir12); /* Measure leakage; */
/* return results to ir12. */
The reverse bias leakage of a diode is measured after applying 40.0 V to the junction.
Also see
None
getstatus
This command returns the operating state of a specified instrument.
Usage
int getstatus(int instr_id, long parameter, double *result);
The instrument identification code
The parameter of query; see Details
The data returned from the instrument; the
command returns one item
Details
If you see the UT_INVLDPRM invalid parameter error returned from the getstatus command, it
indicates that the status item parameter is illegal for this device. The requested status code is invalid
for the selected device.
A list of supported getstatus command values for the parameter parameter for a source-measure
unit (SMU) and a pulse card (VPU) are provided in the following tables.
No status values are provided for measurement-specific conditions.