Section
13: LPT library function reference Model 4200A-SCS Parameter Analyzer
Reference Manual
13-70 4200A-901-01 Rev. C / February 2017
Example
double ir12;
.
.
conpin(2, GND, 0);
conpin(SMU1, 1, 0);
limiti(SMU1, 2.0E-4); /* Limit 1 mA to 200 uA. */
forcev(SMU1, 40.0); /* Apply 40.0 V. */
measi(SMU1, &ir12); /* Measure leakage; */
/* return results to ir12. */
The reverse bias leakage of a diode is measured after applying 40.0 V to the junction.
Also see
None
getstatus
This command returns the operating state of a specified instrument.
Usage
int getstatus(int instr_id, long parameter, double *result);
instr_id
The instrument identification code
parameter
The parameter of query; see Details
result
The data returned from the instrument; the
getstatus
command returns one item
Details
If you see the UT_INVLDPRM invalid parameter error returned from the getstatus command, it
indicates that the status item parameter is illegal for this device. The requested status code is invalid
for the selected device.
A list of supported getstatus command values for the parameter parameter for a source-measure
unit (SMU) and a pulse card (VPU) are provided in the following tables.
No status values are provided for measurement-specific conditions.

Table of Contents

Other manuals for Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200A-SCS and is the answer not in the manual?

Keithley 4200A-SCS Specifications

General IconGeneral
BrandKeithley
Model4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals