Section
9: Keithley External Control Interface (KXCI) Model 4200A-SCS Parameter Analyzer
Reference Manual
9-72 4200A-901-01 Rev. C / February 2017
:CVU:DATA:STATUS?
This command queries the status of the CVU when a test is complete.
Usage
:CVU:DATA:STATUS?
Also see
None
:CVU:DATA:TSTAMP?
This command queries the timestamp of the measurements of the CVU once a test is complete.
Usage
:CVU:DATA:TSTAMP?
Also see
None
Modeless commands
:CVU:CHANNEL chan
This command selects the CVU card on which subsequent CVU commands will act.
Usage
:CVU:CHANNEL chan
chan
CVU card on which KXCI CVU commands act (default
1
)
Details
The majority of systems only have one CVU card. You can use this command to configure multiple
cards.
Also see
None
:CVU:MODEL
This command sets the measurement model for the selected CVU card.
Usage
:CVU:MODEL model
model
The model:
0: Z, theta
1: R + jx
2: Cp, Gp
3: Cs, Rs
4: Cp, D
5
: Cs, D
Also see
None

Table of Contents

Other manuals for Keithley 4200A-SCS