Model 4200A
-SCS Parameter Analyzer Reference Manual Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017 9-73
:CVU:MODE
This command sets User or System mode.
Usage
:CVU:MODE mode
mode
The mode:
0: User Mode
1
: System Mode
Also see
None
:CVU:RESET
This command sends a soft reset to the specified card.
Usage
:CVU:RESET
Details
This command places the card in default state.
Parameters are:
DC voltage: 0 V
AC voltage: 30 mV
Frequency: 100 kHz
Model: R + jx
Range: 1 mA
Also see
None