EasyManuals Logo
Home>Keithley>Measuring Instruments>4200A-SCS

Keithley 4200A-SCS Reference Manual

Keithley 4200A-SCS
1381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1033 background imageLoading...
Page #1033 background image
Section
13: LPT library function reference Model 4200A-SCS Parameter Analyzer
Reference Manual
13-188 4200A-901-01 Rev. C / February 2017
smeasv
This command returns the DC bias voltages used for a sweep.
Usage
int smeasv(int instr_id, double *varray);
instr_id
The instrument identification code of the 4210-CVU:
CVU1
varray
Returned array of DC bias voltages
Details
This command returns the DC bias voltages used in a sweep. The values are returned in an array.
The voltage values are posted to the Clarius Analyze sheet and graph after the test has finished.
You can also use the smeasvRT command to return sourced sweep DC bias voltage values in an
array. It posts the voltage values to Clarius in real time.
Use the measv command to return the DC bias voltage used for a single measurement.
Also see
measv (on page 13-178)
smeasvRT (on page 13-189)

Table of Contents

Other manuals for Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200A-SCS and is the answer not in the manual?

Keithley 4200A-SCS Specifications

General IconGeneral
BrandKeithley
Model4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals