EasyManuals Logo

Keithley 4200A-SCS Reference Manual

Keithley 4200A-SCS
1381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1252 background imageLoading...
Page #1252 background image
Model 4200A
-SCS Parameter Analyzer Reference Manual Appendix G:
Using a Cascade Microtech PA200 Prober
4200A-901-01 Rev. C / February 2017 G-17
5. Manually align pins and pads (POINT 2) and select Finish.
Figure 654: Aligning the wafer: Point 2
Verify wafer alignment
Confirm that the alignment is correct (the alignment procedure is repeated). To check, manually use
the joystick to move the chuck in index moves and confirm that the pins and pads are aligned.
Figure 655: Verify wafer alignment

Table of Contents

Other manuals for Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200A-SCS and is the answer not in the manual?

Keithley 4200A-SCS Specifications

General IconGeneral
BrandKeithley
Model4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals