-SCS Parameter Analyzer Reference Manual Appendix G:
Using a Cascade Microtech PA200 Prober
4200A-901-01 Rev. C / February 2017 G-17
5. Manually align pins and pads (POINT 2) and select Finish.
Figure 654: Aligning the wafer: Point 2
Verify wafer alignment
Confirm that the alignment is correct (the alignment procedure is repeated). To check, manually use
the joystick to move the chuck in index moves and confirm that the pins and pads are aligned.
Figure 655: Verify wafer alignment