EasyManua.ls Logo

Keithley 4200A-SCS - Row-Column or Instrument Card Settings

Keithley 4200A-SCS
1381 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 4200A
-SCS Parameter Analyzer Reference Manual Appendix A: Using switch
matrices
4200A-901-01 Rev. C / February 2017 A-13
Row-column or instrument card settings
You select the scheme for interconnections between the instruments, the switch-matrix rows and
columns, and the test system (prober or test fixture). You can select:
Row-Column: Connect instruments to rows and prober or test fixture to columns.
Instrument Card: Both instruments and prober or test fixture are connected to columns. Matrix
rows are not used.
The row-column setting is the simplest connection scheme. In this scheme, instruments are
connected to the switch-matrix rows. The prober/test fixture pins or the device under test (DUT) are
connected to the switch-matrix columns (see Switch matrix control (on page A-16) and
4200A-SCS
signal paths (on page A-17)).
When you set up a matrix, you also select the sense. You can select:
Local sense: 2-wire conections. Connections are only to instrument FORCE terminals.
Remote sense: 4-wire connections. Connections are to both instrument FORCE and SENSE
terminals.
For more information regarding local and remote sense, refer to Remote sensing (on page 3-29
).
Row-column scheme
The row-column setting is the simplest connection scheme. In this scheme, instruments are
connected to the switch-matrix rows. The prober/test fixture pins or the device under test (DUT) are
connected to the switch-matrix columns (see Switch matrix control (on page A-16) and
4200A-SCS
signal paths (on page A-17)).
Instrument signals can route to prober/test-fixture pins through only one matrix card, as shown in the
following figure. However, the Row-Column scheme limits the number of external instruments. If the
instrumentation requirements exceed eight paths (rows), you must use the instrument card
configuration.
Figure 529: Row-Column, Local Sense Connection Scheme example

Table of Contents

Other manuals for Keithley 4200A-SCS

Related product manuals