EasyManuals Logo

Keithley 4200A-SCS Reference Manual

Keithley 4200A-SCS
1381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1332 background imageLoading...
Page #1332 background image
Model 4200A
-SCS Parameter Analyzer Reference Manual Appendix K:
Using a Signatone CM500 Prober
4200A-901-01 Rev. C / February 2017 K-3
4. Click Addr and verify that the GPIB address matches the GPIB_Address setting in the 4200A-
SCS prober configuration file prbcnfg_CM500.dat at C:\s4200\sys\dat. Note that the
default GPIB address is set to 28.
5. If the address does not match, enter the new GPIB address, then click OK.
Figure 793: Set GPIB Address
Modify the prober configuration file
The default prober configuration file is shown below. As shown, the file is configured for use with a
GPIB communications setup. Use a text editor such as Microsoft
®
Notepad to work with this file if
needed.
On the 4200A-SCS, the configuration file is at C:\s4200\sys\dat\prbcnfg_CM500.dat.

Table of Contents

Other manuals for Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200A-SCS and is the answer not in the manual?

Keithley 4200A-SCS Specifications

General IconGeneral
BrandKeithley
Model4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals