13: LPT library function reference  Model 4200A-SCS Parameter Analyzer 
 
13-190  4200A-901-01 Rev. C / February 2017 
 
Details 
This command makes an impedance measurement on each step of a voltage or frequency sweep. 
The measured values for a sweep are returned in arrays. The measured readings are posted to the 
Clarius Analyze sheet and graph after the test has finished. 
Before calling smeasz, use the forcev command to set the DC bias level, the setfreq command 
to set the AC drive frequency and the setlevel command to set the AC drive voltage. 
Measurement model parameter values 
 
 
 
Impedance (Z) and phase (θ in degrees) 
Parallel capacitance and conductance 
Series capacitance and resistance 
Parallel capacitance and dissipation factor 
Series capacitance and dissipation factor 
You can also use the smeaszRT command to measure and return sweep impedance 
measurements. It posts the measured readings to Clarius in real time. 
To return a single impedance measurement, use measz. 
 
Example 
Programming examples (on page 13-194) 2 through 5 use the smeas command for impedance 
measurements. 
Also see 
forcev (on page 13-175) 
measz (on page 13-178) 
setfreq (on page 13-181) 
setlevel (on page 13-182) 
setmode (on page 13-182) 
smeaszRT (on page 13-191)