Model 4200A
-SCS Parameter Analyzer Reference Manual Appendix D: Using a Model 82 C-
V System
4200A-901-01 Rev. C / February 2017 D-43
The return values from status can be:
0: OK.
-10000 (INVAL_INST_ID): The specified instrument ID does not exist
-10001 (INVAL_PIN_SPEC): An invalid DUT pin number was specified
-10003 (NO_SWITCH_MATRIX): No switch matrix was found
-10004 (NO_MATRIX_CARDS): No matrix cards were found
-10020 (COMP_FILE_ACCESS_ERR): There was an error accessing the specified cable
compensation file
-10021 (COMP_FILE_NOT_EXIST): The specified compensation file does not exist
-10022 (KI590_NOT_IN_KCON): There is no CMTR defined in your system configuration
-10090 (GPIB_ERROR_OCCURRED): A GPIB communications error occurred
-10100 (INVAL_PARAM): An invalid input parameter is specified
If CabCompFile does not exist, or if there is no path specified (null string), the default compensation
parameters are used. When entering the path, be sure to use two \ characters to separate each
directory. For example, if your cable file is in:
C:\calfiles\590cal.dat
You would enter:
C:\\calfiles\\590cal.dat
If you are controlling a switch matrix to route signals using a connection UTM (for example,
“connect”), you do not need connect InputPin and OutputPin. Set these parameters to 0.
Also see
None

Table of Contents

Other manuals for Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200A-SCS and is the answer not in the manual?

Keithley 4200A-SCS Specifications

General IconGeneral
BrandKeithley
Model4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals