EasyManua.ls Logo

Keithley 4200A-SCS

Keithley 4200A-SCS
1381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section
6: Clarius Model 4200A-SCS Parameter Analyzer
Reference Manual
6-168 4200A-901-01 Rev. C / February 2017
Degradation targets
You can enable an output value as a target and assign it a target value (in % change or absolute
value). When all targets for a device are reached, that device is no longer tested. Subsequent cycles
bypass the device tests that reached all its targets. The subsite stops when all targets are reached or
the last subsite cycle is completed.
The testing process for target evaluation is shown in the following flowchart. As a simple example,
assume all the targets for both devices are reached after the first cycle of the subsite test. Following
the flowchart shows that the tests for cycles 2, 3 and 4 are not performed. The subsite test stops.
Figure 330: Target evaluation process (example for two devices, four cycles)
To enable an output value as a target, refer to Export output values to Analyze sheet (on page 6-
176).
The graph that is plotted is degradation versus stress time.

Table of Contents

Other manuals for Keithley 4200A-SCS

Related product manuals