-SCS Parameter Analyzer Reference Manual Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017 9-69
:CVU:DELAY:SWEEP
This command sets the sweep delay for the CVU test on the selected card.
Usage
:CVU:DELAY:SWEEP sweepd
Delay (0 to 999 s)
Details
This command is used when in sweeping mode (which is all sweep types except
:CVU:BIAS:DCV:SAMPLE).
Also see
None
:CVU:SAMPLE:HOLDT
This command sets the hold time for a sampling mode test on the selected card.
Usage
:CVU:SAMPLE:HOLDT holdt
Details
This is only used when executing the :CVU:BIAS:DCV:SAMPLE command.
Also see
:CVU:BIAS:DCV:SAMPLE (on page 9-67)
:CVU:SAMPLE:INTERVAL
This command sets the delay between samples for the selected card.
Usage
:CVU:SAMPLE:INTERVAL interval
Details
This is only used when executing the :CVU:BIAS:DCV:SAMPLE command.
Also see
:CVU:BIAS:DCV:SAMPLE (on page 9-67)