EasyManuals Logo

Keithley 4200A-SCS Reference Manual

Keithley 4200A-SCS
1381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #795 background imageLoading...
Page #795 background image
Section
9: Keithley External Control Interface (KXCI) Model 4200A-SCS Parameter Analyzer
Reference Manual
9-84 4200A-901-01 Rev. C / February 2017
GP
This command queries input or output parameter values, or both, by number for the last user module run in KXCI.
Usage
GP ParameterPosition
GP ParameterPosition NumValues
ParameterPosition
The position of the parameter in the KULT module
NumValues
The number of values in an output array to be returned; see Details
Details
NumValues is only used for an output parameter that is an array. If NumValues is not used, one
value is returned.
Arrays are returned as a list of values separated by semicolons.
The value returned for an input parameter is the given value. The values returned for an output
parameter is the outcome of the test (for example, measured readings).
You can also query a parameter by specifying the name of the parameter in the user module. See
GN (on page 9-83).
Example
GP 5 11
This command queries all test voltages for the Vforce parameter for the user module that was run in the
EX example. Vforce is the fifth parameter in the function.
The following array of test voltages is returned and displayed in the KXCI console:
0; 0.5; 1.0; 1.5; 2.0; 2.5; 3.0; 3.5; 4.0; 4.5; 5.0
Also see
None

Table of Contents

Other manuals for Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 4200A-SCS and is the answer not in the manual?

Keithley 4200A-SCS Specifications

General IconGeneral
BrandKeithley
Model4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals