EasyManua.ls Logo

Keithley 4200A-SCS

Keithley 4200A-SCS
1381 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section
13: LPT library function reference Model 4200A-SCS Parameter Analyzer
Reference Manual
13-14 4200A-901-01 Rev. C / February 2017
To abort a running pulse_exec pulse test, see dev_abort.
devint does the following:
1. Clears all sources by calling devclr.
2. Clears the matrix crosspoints by calling clrcon.
3. Clears the trigger tables by calling clrtrg.
4. Clears the sweep tables by calling clrscn.
5. Resets GPIB instruments by sending the string defined with kibdefint.
6. Resets the active instrument cards.
Instrument cards are reset in the following order:
1. SMU instrument cards
2. CVU instrument cards
3. Pulse instrument cards (4225-PMU or 4220-PGU)
devint is implicitly called by execut and tstdsl. devclr is implicitly called by clrcon.
The SMUs return to the following states:
100 µA and 10 V ranges
Autorange on
Voltage source
0 V DCV bias
The 4210-CVU returns to the following states:
30 mV
RMS
AC signal
0 V DCV bias
100 kHz frequency
Autorange on
Cable length compensation s
et to 0 m
Open/Short/Load compensation disabled

Table of Contents

Other manuals for Keithley 4200A-SCS

Related product manuals